EASYCAL™ User GuideVersion 4.0

Strany 2

10 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideNote:The primary data according to GLP instructions will be transmitted wit

Strany 3

EASYCAL™ User Guide Version 4.0 · Status: June 2008 11EASYCAL™ User GuideIn case the instrument is tested for the first time you will be asked at whi

Strany 4

12 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide3.2 Calibration of volumetric instruments of glass/plasticClick on the gra

Strany 5

EASYCAL™ User Guide Version 4.0 · Status: June 2008 13EASYCAL™ User GuideSubdivision (required entry)Click on the arrow to the right of the entry fie

Strany 6

14 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideEntry of measuring results☞ The entry fields for measuring results have a

Strany 7

EASYCAL™ User Guide Version 4.0 · Status: June 2008 15EASYCAL™ User Guide3.2.2 Entry of error limitsOn the “Entry of tolerance limits” screen you de

Strany 8

16 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide3.2.4 “Save present test” (permanently) and “Intervall entry”4.1 Selectin

Strany 9

EASYCAL™ User Guide Version 4.0 · Status: June 2008 17EASYCAL™ User GuideDefining the filter (Example):Example:☞ Use the mouse to select an option f

Strany 10

18 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide4.7 GraphThe “Curve” button allows you to view a graph of the variouscalibr

Strany 11

EASYCAL™ User Guide Version 4.0 · Status: June 2008 19EASYCAL™ User Guide6.1 Company and LanguageClick on “Please define company and language” to di

Strany 13

20 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideCreate a new test recordClick “New” or “F2” to open a blank entry screen. E

Strany 14

EASYCAL™ User Guide Version 4.0 · Status: June 2008 21EASYCAL™ User Guide6.4 Manage EASYCAL™ dataClick on “Manage Easycal data” to open the following

Strany 15

22 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide“Read test archives”Data which have been saved as an archive can be re-impo

Strany 16

EASYCAL™ User Guide Version 4.0 · Status: June 2008 23EASYCAL™ User GuideChange PasswordYou can change the preferences and the password to protect yo

Strany 17

24 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideCoeffi cient of VariationThe coefficient of variation (CV) is a measure of r

Strany 18

EASYCAL™ User Guide Version 4.0 · Status: June 2008 25EASYCAL™ User GuidePoor accuracy:Hits far off center.Poor reproducibility:Hits widely scattered

Strany 19

26 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide7.2 BasicsThe values obtained by weighing during the gravimetric test are o

Strany 20

EASYCAL™ User Guide Version 4.0 · Status: June 2008 27EASYCAL™ User Guide7.3 Calculations for Glass/PlasticsExample:Serial No: 95 080634Trademark:

Strany 21

28 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideDefinitions in Test Record for Glass/PlasticsDeviation = Maximum deviation

Strany 22

EASYCAL™ User Guide Version 4.0 · Status: June 2008 29EASYCAL™ User GuideTable 1Table 1 provides the factor “Z” for temperatures from 15 °C to 30 °C

Strany 23

EASYCAL™ User Guide Version 4.0 · Status: June 2008 3EASYCAL™ User GuideContents1. Your License Agreement ...

Strany 24

30 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideTable 2If volumetric instruments made of plastic are to be tested, the foll

Strany 25

EASYCAL™ User Guide Version 4.0 · Status: June 2008 31EASYCAL™ User Guide7.4 Calculations for Liquid HandlingExcerpt from ISO 8655/3, Table refers t

Strany 26

32 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide8.1 Printing problemsParts of the record are not printedThis problem occurs

Strany 27

EASYCAL™ User Guide Version 4.0 · Status: June 2008 33EASYCAL™ User Guide8.2 Memory problemsThe MS ACCESS database application, which forms the basis

Strany 28

34 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideAccuracy ... 24Adopt

Strany 30

BRAND GMBH + CO KG · P.O. Box 1155 · 97861 Wertheim · GermanyPhone: +49 9342 808-0 · Fax: +49 9342 808-236 · E-Mail: [email protected]: www.brand.

Strany 31

4 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide1. Your Licence Agreement1.1 Object of Agreement The object of this agree

Strany 32

EASYCAL™ User Guide Version 4.0 · Status: June 2008 5EASYCAL™ User Guide2.1 Program featuresEASYCAL™ is a software system that was developed using Mi

Strany 33

6 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide2.4 Uninstalling the programPrograms that have been specially developed for

Strany 34

EASYCAL™ User Guide Version 4.0 · Status: June 2008 7EASYCAL™ User Guide3. Main Menu3.1 Calibration of Liquid Handling InstrumentsClick on the gray

Strany 35

8 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideEnter the number or the type of the thermometer used in the test. By clickin

Strany 36 - Internet: www.brand.de

EASYCAL™ User Guide Version 4.0 · Status: June 2008 9EASYCAL™ User GuideTest temperature (required entry)The permissible temperature range for measur